Data Foundry

Medical device recalls 2013

Medtronic Intrathecal Catheter — Class I medical device recall Z-1576-2013

Terminated recall by Medtronic Neuromodulation, reported 2013-07-03, distributed nationwide. The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for o

Recall numberZ-1576-2013
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Neuromodulation, Minneapolis, MN, United States
Recall initiated2013-06-03
Classified2013-06-25
Reported by FDA2013-07-03
Terminated2014-04-07
DistributedNationwide (US)
Countries outside the USAE, AT, AU, AW, BE, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, EC, ES, FI, FR, GB, GR, HK, HU, IE, IL, IN, IS, IT, JO, KR, KW, LB, LK, LT, LU, MA, MT, NL, NO, NZ, PA, PL, PT, QA, RO, SA, SE, SG, SI, SK, TH, TR, TT, UY, VE, ZA
Quantity115,722 total
Reason classesdevice malfunction
Model numbers8731SC

Product

Medtronic Intrathecal Catheter, model 8731SC. Contents: 38.1-cm spinal segment with length markings and guide wire, 66-cm pump segment, 15 T-gauge introducer needle, Transparent Strain-relief sleeves, Opaque Strain-relief sleeves, and V-wing anchors. Contents of inner package are STERILE. The implantable Medtronic Model 8731SC Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and a Model 8731SC catheter. The catheter connects to the pump at the catheter port.

Reason for recall

The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.