STA Compact Automated Multi-Parametric Analyzer — Class II medical device recall Z-1576-2023
Ongoing recall by Diagnostica Stago, Inc., reported 2023-05-17, distributed nationwide. An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent s
| Recall number | Z-1576-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Diagnostica Stago, Inc., Parsippany, NJ, United States |
| Recall initiated | 2023-03-22 |
| Classified | 2023-05-10 |
| Reported by FDA | 2023-05-17 |
| Distributed | Nationwide (US) |
| Quantity | 2 analyers |
| Reason classes | software |
| Serial numbers | 7096527, 7116678 |
Product
STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612
Reason for recall
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).