Data Foundry

Medical device recalls 2026

Battery Charging Station — Class III medical device recall Z-1587-2026

Ongoing recall by Datascope Corp., reported 2026-03-25, distributed nationwide. The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion b

Recall numberZ-1587-2026
Product typeMedical device
ClassificationClass III
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDatascope Corp., Mahwah, NJ, United States
Recall initiated2026-02-06
Classified2026-03-19
Reported by FDA2026-03-25
DistributedNationwide (US)
Countries outside the USAU, HK, IL, JP, NO
UPC / GTIN / UDI-DI10607567111964
Serial numbers200017647EAI, 200019347EAI, 200019447EAI, 200019847EAI, 210017525EAI, 210017925EAI, 210018025EAI, 210018425EAI, 210018525EAI, 210018625EAI, 210019025EAI, 210019125EAI, 210019225EAI, 210019725EAI, 210020044EAI, 210020144EAI, 210020244EAI, 210020544EAI, 210020944EAI, 210021044EAI, 210021144EAI, 210021344EAI, 210022144EAI, 210022344EAI, 210022544EAI and 3 more
Model numbers0998-00-0802

Product

Battery Charging Station; Model: 0998-00-0802;

Reason for recall

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.