Data Foundry

Medical device recalls 2026

Handy Solutions Neck & Shoulders Heating Pad — Class I medical device recall Z-1588-2026

Ongoing recall by Navajo Manufacturing Company, reported 2026-04-01, distributed nationwide. If heating pad is folded while in use, the power density increases in the folded area causing excessive temper

Recall numberZ-1588-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmNavajo Manufacturing Company, Denver, CO, United States
Recall initiated2026-02-18
Classified2026-03-20
Reported by FDA2026-04-01
DistributedNationwide (US)
Quantity28,457
UPC / GTIN / UDI-DI00024291256076

Product

Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd

Reason for recall

If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1588-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.