Handy Solutions Neck & Shoulders Heating Pad — Class I medical device recall Z-1588-2026
Ongoing recall by Navajo Manufacturing Company, reported 2026-04-01, distributed nationwide. If heating pad is folded while in use, the power density increases in the folded area causing excessive temper
| Recall number | Z-1588-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Navajo Manufacturing Company, Denver, CO, United States |
| Recall initiated | 2026-02-18 |
| Classified | 2026-03-20 |
| Reported by FDA | 2026-04-01 |
| Distributed | Nationwide (US) |
| Quantity | 28,457 |
| UPC / GTIN / UDI-DI | 00024291256076 |
Product
Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
Reason for recall
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1588-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.