Data Foundry

Medical device recalls 2024

Identity Imprint PS Tibial Tray Size 4: Lot 540287 — Class II medical device recall Z-1591-2024

Ongoing recall by Conformis, Inc., reported 2024-04-24, distributed nationwide. Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540

Recall numberZ-1591-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmConformis, Inc., Billerica, MA, United States
Recall initiated2024-03-11
Classified2024-04-17
Reported by FDA2024-04-24
DistributedNationwide (US)
Quantity3 units
Lot numbers540287
Model numbersTPS-301-T04S

Product

Identity Imprint PS Tibial Tray Size 4: Lot 540287

Reason for recall

Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1591-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.