Data Foundry

Medical device recalls 2024

Siemens Artis Models: Interventional Fluoroscopic X-Ray… — Class II medical device recall Z-1598-2024

Ongoing recall by Siemens Medical Solutions USA, Inc, reported 2024-04-24, distributed nationwide. Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the

Recall numberZ-1598-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2024-03-21
Classified2024-04-18
Reported by FDA2024-04-24
DistributedNationwide (US)
Quantity583 units
Reason classesdevice malfunction
UPC / GTIN / UDI-DI04056869009971, 04056869009988, 04056869009995, 04056869010007, 04056869010014, 04056869010021, 04056869010038, 04056869010045, 04056869010052, 04056869010069, 04056869010076, 04056869010083, 04056869010090, 04056869010106

Product

Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143

Reason for recall

Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1598-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.