Alinity m System — Class II medical device recall Z-1602-2024
Ongoing recall by Abbott Molecular, Inc., reported 2024-05-01, distributed nationwide. The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention
| Recall number | Z-1602-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Molecular, Inc., Des Plaines, IL, United States |
| Recall initiated | 2024-03-20 |
| Classified | 2024-04-19 |
| Reported by FDA | 2024-05-01 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, EE, EG, ES, FI, FR, GB, GR, HK, HR, IE, IL, IN, IT, JP, KE, KR, LU, LV, MA, MX, MY, NG, NL, NO, NZ, PA, PK, PL, PT, PY, RO, RU, SA, SE, SG, SI, SV, TH, TR, TW, VN, ZA |
| Quantity | 1,012 units |
| UPC / GTIN / UDI-DI | 00884999048034 |
| Model numbers | 08N53-002 |
Product
Alinity m System, Part Number: 08N53-002
Reason for recall
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.