I — Class II medical device recall Z-1604-2026
Ongoing recall by I.T.S. GmbH, reported 2026-04-01, distributed nationwide. Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions t
| Recall number | Z-1604-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | I.T.S. GmbH, Lasnitzhohe, Austria |
| Recall initiated | 2026-01-23 |
| Classified | 2026-03-20 |
| Reported by FDA | 2026-04-01 |
| Distributed | Nationwide (US) |
| Quantity | 120 units |
| Lot numbers | 21231-6, 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4, 37302-18, 37302-20, 643/022310, 643/032308, 643/0630, 643/0657, 643/102402 |
Product
I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.
Reason for recall
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).