Data Foundry

Medical device recalls 2015

Dimension EXL with LM — Class II medical device recall Z-1615-2015

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2015-05-20. Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a

Recall numberZ-1615-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2015-04-02
Classified2015-05-14
Reported by FDA2015-05-20
Terminated2016-08-17
Countries outside the USJP
Quantity17,801
Model numbersD045912, D140066, D193440

Product

Dimension EXL with LM, Dimension EXL 200, Dimension RxL, Dimension RxL HM, Dimension RxL Max, Dimension RxL Max HM, Dimension Xpand, Dimension Xpand HM, Dimension Xpand Plus, Dimension Xpand Plus HM. The Dimension EXL clinical chemistry system is a discrete, random-access, microprocessor-controlled, integrated instrument/chemistry system that measures a variety of analytes, including enzyme activities, in body fluids for in vitro diagnostic use.

Reason for recall

Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical hazard. The likelihood of potential clinical impact or injury due to this issue is extremely low.