MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433 — Class II medical device recall Z-1622-2013

Terminated recall by Maquet Cardiovascular Us Sales, Llc, reported 2013-07-10, distributed in AZ, GA, ND, SC, TN, TX. MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make

Recall numberZ-1622-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMaquet Cardiovascular Us Sales, Llc, Wayne, NJ, United States
Recall initiated2013-04-26
Classified2013-07-01
Reported by FDA2013-07-10
Terminated2014-04-29
DistributedAZ, GA, ND, SC, TN, TX
Countries outside the USAE, AT, BR, CA, CH, CN, DE, ES, FR, GB, IT, JP, RS, SG, TH, TR, ZA

Product

MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment.

Reason for recall

MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1622-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.