Homocysteine test system — Class II medical device recall Z-1622-2015
Terminated recall by Roche Diagnostics Operations, Inc., reported 2015-05-27, distributed nationwide. Customers complained about under-recovery of non-Roche controls and discrepant low patient results with certai
| Recall number | Z-1622-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Operations, Inc., Indianapolis, IN, United States |
| Recall initiated | 2015-03-23 |
| Classified | 2015-05-15 |
| Reported by FDA | 2015-05-27 |
| Terminated | 2016-01-19 |
| Distributed | Nationwide (US) |
| Quantity | 2,558 units |
| Lot numbers | 69781101 |
Product
Homocysteine test system - Homocysteine Reagent. 05385415190; Homocysteine 100 Tests, cobas c, COBAS INTEGRA Homocysteine Enzymatic Assay is an in vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems.
Reason for recall
Customers complained about under-recovery of non-Roche controls and discrepant low patient results with certain Homocysteine reagent lots. This negative bias could, in the worst case, lead to inaccurately low Homocysteine results. An elevated level of Homocysteine is considered an important risk factor in the assessment of peripheral vascular disease. Falsely low values could lead to a delay of di