Data Foundry

Medical device recalls 2020

Philips V60 Ventilators with Power Management PCBA part number 1055906 — Class II medical device recall Z-1622-2020

Ongoing recall by Respironics California, LLC, reported 2020-04-08, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blowe

Recall numberZ-1622-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRespironics California, LLC, Carlsbad, CA, United States
Recall initiated2020-03-05
Classified2020-03-31
Reported by FDA2020-04-08
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY
Countries outside the USAE, AR, AT, AU, BE, BH, BO, BR, CA, CH, CL, CN, CZ, DE, DK, DO, EC, EG, ES, FI, FR, GB, HK, HR, ID, IE, IL, IN, IR, IS, IT, KE, KR, KW, LB, LU, MA, MT, MX, MY, NG, NL, NO, NZ, OM, PA, PE, PH, PL, PT, PY, QA, SA, SE, SG, SI, TH, TR, TW, VE, VN, ZA
Quantity23,581
Serial numbers100002908, 100027093, 100027096, 100085389, 100085399, 100109746, 100109868, 201003390, 201003392, 201010952
Model numbers1055906

Product

Philips V60 Ventilators with Power Management PCBA part number 1055906

Reason for recall

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1622-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.