Philips V60 Ventilators with Power Management PCBA part number 1055906 — Class II medical device recall Z-1622-2020
Ongoing recall by Respironics California, LLC, reported 2020-04-08, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blowe
| Recall number | Z-1622-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Respironics California, LLC, Carlsbad, CA, United States |
| Recall initiated | 2020-03-05 |
| Classified | 2020-03-31 |
| Reported by FDA | 2020-04-08 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY |
| Countries outside the US | AE, AR, AT, AU, BE, BH, BO, BR, CA, CH, CL, CN, CZ, DE, DK, DO, EC, EG, ES, FI, FR, GB, HK, HR, ID, IE, IL, IN, IR, IS, IT, KE, KR, KW, LB, LU, MA, MT, MX, MY, NG, NL, NO, NZ, OM, PA, PE, PH, PL, PT, PY, QA, SA, SE, SG, SI, TH, TR, TW, VE, VN, ZA |
| Quantity | 23,581 |
| Serial numbers | 100002908, 100027093, 100027096, 100085389, 100085399, 100109746, 100109868, 201003390, 201003392, 201010952 |
| Model numbers | 1055906 |
Product
Philips V60 Ventilators with Power Management PCBA part number 1055906
Reason for recall
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia