Data Foundry

Medical device recalls 2015

IMMULITE/IMMULITE 1000 Systems TBG — Class II medical device recall Z-1629-2015

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2015-05-27. IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parameter of <1.80.

Recall numberZ-1629-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2015-04-09
Classified2015-05-19
Reported by FDA2015-05-27
Terminated2017-03-22
Countries outside the USCN, DE, DK, ES, GB
Quantity155 kits
Lot numbers0257, 10381643
Model numbersLKTB1

Product

IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement of Thyroxine-Binding Globulin (TBG) in serum, as an aid in the diagnosis and treatment of thyroid diseases.

Reason for recall

IMMULITE/IMMULITE 1000 TBG catalog LKTB lot 257 is failing the adjustment slope instrument parameter of <1.80. The resulting slope of the kit is >1.80 resulting in the potential to produce depressed TBG patient values.