EPIQ 5 Ultrasound System versions 1 — Class II medical device recall Z-1631-2015

Terminated recall by Philips Ultrasound, Inc., reported 2015-05-27. When EPIQ 5 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/o

Recall numberZ-1631-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Ultrasound, Inc., Bothell, WA, United States
Recall initiated2015-04-14
Classified2015-05-19
Reported by FDA2015-05-27
Terminated2016-08-08
Model numbers795204, 795205, 989605408541

Product

EPIQ 5 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W.Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3) 453561800581 (1.3.1) 453561805181 (1.3.2) Diagnostic ultrasound imaging and fluid flow analysis.

Reason for recall

When EPIQ 5 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).