Data Foundry

Medical device recalls 2016

CinchLock Flex Knotless Anchor with Inserter — Class II medical device recall Z-1635-2016

Terminated recall by Stryker Corporation, reported 2016-05-18, distributed nationwide. Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor wi

Recall numberZ-1635-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Corporation, San Jose, CA, United States
Recall initiated2016-04-12
Classified2016-05-10
Reported by FDA2016-05-18
Terminated2016-12-12
DistributedNationwide (US)
Countries outside the USAU, CH, ES, NL, SE
Quantity3,100 devices
Serial numbers15122202, 16011301, 16020802
Model numbers02462, 02463, 02643

Product

CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex Knotless Anchor (CAT02463) are non-degradable implant devices intended for the fixation of soft tissue to bone. Used for repair of hip, shoulder, foot and ankle, elbow, wrist and hand and knee.

Reason for recall

Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of an internal wire breaking upon deployment of the anchor.