MiniMed" 780G — Class II medical device recall Z-1635-2026
Ongoing recall by Medtronic MiniMed, Inc., reported 2026-04-01, distributed nationwide. Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a
| Recall number | Z-1635-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic MiniMed, Inc., Northridge, CA, United States |
| Recall initiated | 2026-01-30 |
| Classified | 2026-03-24 |
| Reported by FDA | 2026-04-01 |
| Distributed | Nationwide (US) |
| Countries outside the US | IL |
| Quantity | 26,851 pumps |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00199150003686, 00763000384272, 00763000414344, 00763000439859, 00763000439866, 00763000545604, 00763000578640, 00763000586195, 00763000640095, 00763000758547, 00763000819378, 00763000819385, 00763000819392, 00763000865665, 00763000924089, 00763000924102, 00763000924119 |
| Serial numbers | 10783H, 1911H, 3983804H, 763000758547, 96H, G3748686H, NG2371419H, NG2372069H, NG2372262H, NG2372296H, NG2372322H, NG2372478H, NG2372552H, NG2372569H, NG2372727H, NG2372822H, NG2372829H, NG2372873H, NG2372896H, NG2372897H, NG2373314H, NG2373351H, NG2374772H, NG2376566H, NG2376619H and 475 more |
| Model numbers | DEMMT-1884, MMT-1884, MMT-1884A, MMT-1884AL, MMT-1884G, MMT-1884L, MMT-1886 |
Product
MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
Reason for recall
Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1635-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.