Tandem pumps are battery-operated infusion pumps capable of both… — Class II medical device recall Z-1636-2026

Ongoing recall by Tandem Diabetes Care, Inc., reported 2026-04-01. The Czech language user guide contained multiple translation errors. The most significant error was that it in

Recall numberZ-1636-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTandem Diabetes Care, Inc., San Diego, CA, United States
Recall initiated2026-01-30
Classified2026-03-25
Reported by FDA2026-04-01
Countries outside the USCZ, SK
Quantity1,380 units

Product

Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).

Reason for recall

The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.