Data Foundry

Medical device recalls 2020

BD WAVELINQ 4F EndoAVF System — Class II medical device recall Z-1637-2020

Terminated recall by Bard Peripheral Vascular Inc, reported 2020-04-08, distributed nationwide. The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at r

Recall numberZ-1637-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBard Peripheral Vascular Inc, Tempe, AZ, United States
Recall initiated2019-04-12
Classified2020-04-02
Reported by FDA2020-04-08
Terminated2021-07-28
DistributedNationwide (US)
Quantity150 units
UPC / GTIN / UDI-DI00801741182754
Lot numbers0053
Serial numbersS0053
Model numbersW04200

Product

BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.

Reason for recall

The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at risk of experiencing a magnetic deficiency, which may result in the magnets failing to attract to one another. This may result in procedural delay in obtaining or creating a functional fistula and need for additional contrast to perform the necessary fluoroscopy. This can pose an incremental risk of harm to a diabetic patient's kidney function.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1637-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.