Data Foundry

Medical device recalls 2026

Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord — Class II medical device recall Z-1639-2026

Ongoing recall by Olympus Corporation of the Americas, reported 2026-04-01, distributed nationwide. Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately valid

Recall numberZ-1639-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOlympus Corporation of the Americas, Center Valley, PA, United States
Recall initiated2026-02-25
Classified2026-03-25
Reported by FDA2026-04-01
DistributedNationwide (US)
Countries outside the USAU, CA, CN, JP, KR, MX, SG
Quantity164 units
UPC / GTIN / UDI-DI00821925035898
Model numbers3006

Product

Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.

Reason for recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1639-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.