1) VITROS 5 — Class II medical device recall Z-1647-2013
Terminated recall by Ortho-Clinical Diagnostics, reported 2013-07-10, distributed nationwide. There is evidence that an electrical short within the systems connector has caused an accumulation of heat whi
| Recall number | Z-1647-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2013-05-06 |
| Classified | 2013-07-01 |
| Reported by FDA | 2013-07-10 |
| Terminated | 2018-05-16 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE |
Product
1) VITROS 5,1 FS Chemistry System (product code 6801375) and 2) VITROS 5,1 FS Chemistry System Refurbished (product code 6801890) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Reason for recall
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.