Data Foundry

Medical device recalls 2023

Philips Azurion System with software versions: R2 — Class II medical device recall Z-1647-2023

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2023-06-07, distributed nationwide. Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available

Recall numberZ-1647-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands
Recall initiated2023-03-21
Classified2023-05-26
Reported by FDA2023-06-07
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BD, BE, BG, BH, BO, BR, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VN, ZA
UPC / GTIN / UDI-DI00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272
Model numbers3M12, 722063, 722221

Product

Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224

Reason for recall

Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1647-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.