Data Foundry

Medical device recalls 2013

ORTHO¿ Assay Software — Class II medical device recall Z-1651-2013

Terminated recall by Ortho-Clinical Diagnostics, reported 2013-07-10, distributed nationwide. There is evidence that an electrical short within the systems connector has caused an accumulation of heat whi

Recall numberZ-1651-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2013-05-06
Classified2013-07-01
Reported by FDA2013-07-10
Terminated2018-05-16
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE
Quantity31 units
Serial numbers0811F06216
Model numbers6842840

Product

ORTHO¿ Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO¿ Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing.

Reason for recall

There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.