Data Foundry

Medical device recalls 2016

ADVIA CHEMISTRY ENZYMATIC CREATININE_2 — Class II medical device recall Z-1654-2016

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2016-05-18, distributed nationwide. Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interferen

Recall numberZ-1654-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2016-03-17
Classified2016-05-11
Reported by FDA2016-05-18
Terminated2017-09-25
DistributedNationwide (US)
Quantity137,422 units
Model numbers74330

Product

ADVIA CHEMISTRY ENZYMATIC CREATININE_2

Reason for recall

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1654-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.