Data Foundry

Medical device recalls 2023

Teleflex HORIZON Microclip Titanium Ligating Clips w/tape — Class II medical device recall Z-1663-2023

Ongoing recall by TELEFLEX LLC, reported 2023-06-07. Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected

Recall numberZ-1663-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTELEFLEX LLC, Morrisville, NC, United States
Recall initiated2023-04-28
Classified2023-05-30
Reported by FDA2023-06-07
Quantity146,606 units
UPC / GTIN / UDI-DI24026704696394
Lot numbers73A2200857, 73A2300018, 73A2300176, 73A2300326, 73A2300520, 73A2300778, 73D2200216, 73D2200559, 73E2200634, 73E2200874, 73E2201099, 73F2200193, 73F2200200, 73F2200482, 73F2200711, 73G2200168, 73G2200345, 73G2200560, 73G2200754, 73H2200115, 73H2200294, 73H2200433, 73H2200453, 73J2200747, 73J2200763, 73K2200014, 73K2200095, 73K2200101, 73K2200299, 73K2200566, 73K2200713, 73L2200064, 73L2200066
Model numbers005200

Product

Teleflex HORIZON Microclip Titanium Ligating Clips w/tape, REF 005200; for use in general surgical procedures that require vessel or tissue ligation

Reason for recall

Teleflex LLC is initiating this voluntary field corrective action because the clip cartridges for the affected product is missing a key feature that restricts the application of the clip to the applier. With this step feature missing on the cartridge, the clip is mounted in a more variable position, allowing the clips to load farther back in the clip applier jaw grooves. This may potentially result in inappropriate or inadequate clip application to vascular structures.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1663-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.