Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount — Class II medical device recall Z-1663-2025
Ongoing recall by IHB OPERATIONS B.V., reported 2025-04-30, distributed nationwide. The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engag
| Recall number | Z-1663-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | IHB OPERATIONS B.V., Zwolle, N/A, Netherlands |
| Recall initiated | 2025-03-27 |
| Classified | 2025-04-23 |
| Reported by FDA | 2025-04-30 |
| Distributed | Nationwide (US) |
| Quantity | 112 units |
| UPC / GTIN / UDI-DI | 10841736116071 |
Product
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Reason for recall
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.