Data Foundry

Medical device recalls 2013

Medtronic MiniMed Paradigm Insulin Infusion Sets The Paradigm… — Class I medical device recall Z-1668-2013

Terminated recall by Medtronic MiniMed, reported 2013-07-17, distributed nationwide. Medtronic is recalling the Medtronic MiniMed Paradigm Medtronic is recalling the Medtronic MiniMed Paradigm In

Recall numberZ-1668-2013
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic MiniMed, Northridge, CA, United States
Recall initiated2013-06-07
Classified2013-07-09
Reported by FDA2013-07-17
Terminated2015-06-10
DistributedNationwide (US)
Countries outside the USAE, AR, AU, BH, BM, BR, BS, CA, CH, CL, CN, CR, CY, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HR, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, LV, MA, MX, MY, NL, NO, NZ, OM, PA, PE, PL, PY, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TR, TT, TW, UA, UY, VE, ZA
Quantity11,000,000 units
Reason classesdevice malfunction
Model numbersMMT-312L, MMT-312S, MMT-317, MMT-318, MMT-324, MMT-325, MMT-368, MMT-377, MMT-378, MMT-381, MMT-382, MMT-383, MMT-384, MMT-386, MMT-387, MMT-394, MMT-396, MMT-397, MMT-398, MMT-399, MMT-862, MMT-864, MMT-866, MMT-874, MMT-876 and 12 more

Product

Medtronic MiniMed Paradigm Insulin Infusion Sets The Paradigm infusion sets are intended for use with Paradigm insulin infusion pumps.

Reason for recall

Medtronic is recalling the Medtronic MiniMed Paradigm Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion sets because of a potential safety issue that can occur if insulin or other fluids come in contact with the inside of Medtronic Paradigm infusion sets. Under certain conditions, the infusion set may malfunction and deliver an incorrect volume of insulin.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1668-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.