Data Foundry

Medical device recalls 2016

ADVIA Chemistry Triglyceride_2 — Class II medical device recall Z-1670-2016

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2016-05-25, distributed in CA, CO, CT, FL, KS, MI, MN, NC…. Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) do

Recall numberZ-1670-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2016-03-07
Classified2016-05-13
Reported by FDA2016-05-25
Terminated2017-12-01
DistributedCA, CO, CT, FL, KS, MI, MN, NC, NM, NY, OK, OR, TN, TX, UT, WA
Countries outside the USAT, AU, BE, BG, BR, CZ, DE, DK, EE, ES, FR, GB, IN, IT, LV, NO, NZ, PT, RO, SE, SI, SK, TH, TR
Reason classespotency
Lot numbers348297, 359932
Model numbers10697575

Product

ADVIA Chemistry Triglyceride_2, concentrated; TRIG_c; Catalog # SMN # 10697575. For in vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma on ADVIA Chemistry systems.

Reason for recall

Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions For Use (IFU) linearity claims at the upper limit of the assay range as it approaches end of shelf life for lots # 348297 and 359932.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1670-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.