ARCHITECT Intact PTH Controls consists of 3 x 1 Bottle — Class II medical device recall Z-1672-2014
Terminated recall by Abbott Laboratories, reported 2014-06-04, distributed nationwide. Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate
| Recall number | Z-1672-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Laboratories, Abbott Park, IL, United States |
| Recall initiated | 2014-02-12 |
| Classified | 2014-05-29 |
| Reported by FDA | 2014-06-04 |
| Terminated | 2015-05-22 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, BB, BD, BE, BG, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JM, JO, JP, KE, KR, KW, KY, LB, LK, LT, LU, LV, MA, MX, MY, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA |
| Reason classes | potency |
| Lot numbers | 00713A000, 00713I000, 01012F000, 01412K000, 01813C000, 01813K000, 02613E000 |
| Expiration dates | 2013-05-30, 2013-11-12, 2014-01-08, 2014-03-14, 2014-05-09, 2014-09-05, 2014-11-12 |
Product
ARCHITECT Intact PTH Controls consists of 3 x 1 Bottle (8.0 mL each) of PTH (synthetic peptide) in Bis Tris Propane buffer with protein (bovine) stabilizer. Preservatives: sodium azide and ProClin 300. Product Usage: The ARCHITECT Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System.
Reason for recall
Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate falsely elevated results on patient samples. Results generated with impacted lots may demonstrate a positive shift relative to those generated with previous calibrator and/or control lots. The issue may also impact your established ARCHITECT Intact PTH reference ranges. Abbott ARCHITECT In