Data Foundry

Medical device recalls 2025

Stratus CS Acute Care cTNI TestPak — Class II medical device recall Z-1673-2025

Ongoing recall by Siemens Healthcare Diagnostics Inc, reported 2025-05-07, distributed nationwide. Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for ca

Recall numberZ-1673-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics Inc, Norwood, MA, United States
Recall initiated2025-03-31
Classified2025-04-25
Reported by FDA2025-05-07
DistributedNationwide (US)
Countries outside the USAT, CA, CH, CL, DE, ES, FR, IQ, IT, MY, RS, SA, SE, SI, VN
Quantity367 units
Lot numbers234337002

Product

Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.

Reason for recall

Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.