Data Foundry

Medical device recalls 2013

Four Lead TUR irrigation Set — Class II medical device recall Z-1682-2013

Terminated recall by Baxter Healthcare Corp., reported 2013-07-17, distributed nationwide. A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinhole

Recall numberZ-1682-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corp., Deerfield, IL, United States
Recall initiated2013-05-15
Classified2013-07-10
Reported by FDA2013-07-17
Terminated2015-06-16
DistributedNationwide (US)
Quantity356,728 units
Reason classessterility

Product

Four Lead TUR irrigation Set, for use with UROMATIC Plastic Containers, Product Code 2C4013 - Approximate Length 85" (2.2. m) Usage: Baxter irrigation sets are intended for the delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cytoscopy, transurethral resection and arthroscopic procedures. the sets are intended for use with Baxter irrigating solutions in Uromatic or Arthromatic plastic containers.

Reason for recall

A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1682-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.