Data Foundry

Medical device recalls 2022

Product Description/Item: ChemoLock Vial Spike — Class II medical device recall Z-1682-2022

Ongoing recall by ICU Medical, Inc., reported 2022-09-14, distributed nationwide. Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, d

Recall numberZ-1682-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmICU Medical, Inc., San Clemente, CA, United States
Recall initiated2022-06-27
Classified2022-09-07
Reported by FDA2022-09-14
DistributedNationwide (US)
Countries outside the USAE, AU, CA, ES, JP, SA, SI
Quantity358,975
Lot numbers011-CL-51/5850505, 011-CL-80S/5843950, 011-CL4155/5816324, 034-CL-70/5816321, 034-CL-72/5816344, 034-CL2100/5842230, 5829878, 5842209, 5842210, 5850503, 5851021, 5851022, 5872831, 5872840, 5885020, 5910255, 5910258, 5910262, CL-10/5850516, CL-17/5816314, CL3950/5845117, CL3951/5842199, CL3967/5829901, CL4118/5825274, KL-BS001U3/5851020, KL-VA201U3/5819364, KL-VA202U3/5819370, KL-VA321U3/5915973, KL-VA322U3/5920744
Model numbers011-CL-80S/5843950

Product

Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S; ChemoLock Bag Spike/CL-10; ChemoLock w/Mini Bag Spike/CL-17; 8 IN (20cm) APPX 1.0ml, EXT SET, MicroClave Clear, ChemoLock PORT, YCONN/CL3950; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger/CL3951; 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock Port, Clamp, Graduated Adapter/CL3967; 6 IN(15cm)APPX 0.82ml,EXT SET,MicroClave Clear,ChemoLock PORT,YCONN, ROTATING LUER/CL4118; ChemoLock Kit w/ 5" (13 cm) Bag Spike Adapter w/ChemoLock w/Red Cap, Vented Cap, ChemoLock Port Bag Spike/CL4158; ChemosafeLock Bag Spike/KL-BS001U3; ChemosafeLock Vial Adapter/KL-VA201U3; ChemosafeLock Vial Adapter/KL-VA202U3; ChemosafeLock Vial Adapter/KL-VA321U3; ChemosafeLock Vial Adapter/KL-VA322U3; ChemoLock Vented Vial Spike/011-CL-51; Transfer Set, PUR Yellow w/ChemoLock Port, Spiros/011-CL4155; Puerto ChemoLock/034-CL2100; ChemoLock, Punzon Univ. Vial c/ Filtro, fijacion/034-CL-70; ChemoLock, Punzon Vial Tritan c/ Filtro, 13mm/034-CL-72

Reason for recall

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1682-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.