Data Foundry

Medical device recalls 2015

Neusoft NeuViz 64 Multi-slice CT Scanner System — Class II medical device recall Z-1693-2015

Terminated recall by Neusoft Medical System Co, reported 2015-06-24, distributed in IL, MA, PR, SC, TX. The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it.

Recall numberZ-1693-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmNeusoft Medical System Co, Houston, TX, United States
Recall initiated2014-03-08
Classified2015-06-16
Reported by FDA2015-06-24
Terminated2016-12-05
DistributedIL, MA, PR, SC, TX
Countries outside the USAE, BR, CN, EC, EG, LB, MY, TN, ZA
Quantity5 installed
Reason classesspecification failure

Product

Neusoft NeuViz 64 Multi-slice CT Scanner System

Reason for recall

The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1693-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.