Neusoft NeuViz 64 Multi-slice CT Scanner System — Class II medical device recall Z-1693-2015
Terminated recall by Neusoft Medical System Co, reported 2015-06-24, distributed in IL, MA, PR, SC, TX. The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it.
| Recall number | Z-1693-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Neusoft Medical System Co, Houston, TX, United States |
| Recall initiated | 2014-03-08 |
| Classified | 2015-06-16 |
| Reported by FDA | 2015-06-24 |
| Terminated | 2016-12-05 |
| Distributed | IL, MA, PR, SC, TX |
| Countries outside the US | AE, BR, CN, EC, EG, LB, MY, TN, ZA |
| Quantity | 5 installed |
| Reason classes | specification failure |
Product
Neusoft NeuViz 64 Multi-slice CT Scanner System
Reason for recall
The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image