Data Foundry

Medical device recalls 2017

Proteus 235 — Class II medical device recall Z-1695-2017

Terminated recall by Ion Beam Applications S.A., reported 2017-04-05, distributed in FL, MA, TN, VA. The Proton Therapy System can allow gantry rotation even if the snout is not properly locked.

Recall numberZ-1695-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIon Beam Applications S.A., Louvain La Neuve, Belgium
Recall initiated2017-03-09
Classified2017-03-29
Reported by FDA2017-04-05
Terminated2019-03-26
DistributedFL, MA, TN, VA
Countries outside the USCN, DE, FR, KR
Quantity8 worldwide
Serial numbersPAT.000, PAT.001, PAT.003, PAT.006, PAT.108, PAT.111, SAT.119, SAT.120

Product

Proteus 235

Reason for recall

The Proton Therapy System can allow gantry rotation even if the snout is not properly locked.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1695-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.