Data Foundry

Medical device recalls 2026

Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product… — Class II medical device recall Z-1695-2026

Ongoing recall by Olympus Corporation of the Americas, reported 2026-04-08, distributed nationwide. Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in s

Recall numberZ-1695-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOlympus Corporation of the Americas, Center Valley, PA, United States
Recall initiated2026-02-27
Classified2026-03-30
Reported by FDA2026-04-08
DistributedNationwide (US)
Countries outside the USAU, CA, CL, HK, SG
Quantity32 units
UPC / GTIN / UDI-DI00821925044135, 00821925044593
Serial numbersMDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164 and 16 more

Product

Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No

Reason for recall

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.