Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product… — Class II medical device recall Z-1695-2026
Ongoing recall by Olympus Corporation of the Americas, reported 2026-04-08, distributed nationwide. Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in s
| Recall number | Z-1695-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Olympus Corporation of the Americas, Center Valley, PA, United States |
| Recall initiated | 2026-02-27 |
| Classified | 2026-03-30 |
| Reported by FDA | 2026-04-08 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CL, HK, SG |
| Quantity | 32 units |
| UPC / GTIN / UDI-DI | 00821925044135, 00821925044593 |
| Serial numbers | MDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164 and 16 more |
Product
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Reason for recall
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.