Data Foundry

Medical device recalls 2020

STA — Class II medical device recall Z-1699-2020

Terminated recall by Diagnostica Stago, Inc., reported 2020-04-22, distributed nationwide. A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clot

Recall numberZ-1699-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDiagnostica Stago, Inc., Parsippany, NJ, United States
Recall initiated2020-03-04
Classified2020-04-10
Reported by FDA2020-04-22
Terminated2021-11-10
DistributedNationwide (US)
Quantity3,225 units

Product

STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979

Reason for recall

A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1699-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.