Data Foundry

Medical device recalls 2018

BARD¿ BrachySource¿ Iodine125 radioactive seeds consist of a welded… — Class II medical device recall Z-1702-2018

Terminated recall by Bard Brachytherapy, Inc., reported 2018-05-16, distributed in CA, GA. Bard Medical Division is recalling the BARD¿ BrachySource¿ Iodine125 Radioactive Seeds in Mick¿ Applicators an

Recall numberZ-1702-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBard Brachytherapy, Inc., Carol Stream, IL, United States
Recall initiated2018-04-10
Classified2018-05-07
Reported by FDA2018-05-16
Terminated2020-06-05
DistributedCA, GA
Reason classessterility
Lot numbers23498276, BBCP0079, BBCQ0026, BBCQ0050

Product

BARD¿ BrachySource¿ Iodine125 radioactive seeds consist of a welded titanium capsule containing the I125 absorbed onto a nickel / copper-coated, gold-cored aluminum wire (reference Figure 1 below). The I125 has a half-life of 59.6 days and decays by electron capture with the emission of characteristic photons and Auger electrons. The principal photon emissions are 27.4 and 31 keV x-rays and 35.5 keV gamma. The titanium wall of the BrachySource¿ Seed Implants absorbs the electrons.

Reason for recall

Bard Medical Division is recalling the BARD¿ BrachySource¿ Iodine125 Radioactive Seeds in Mick¿ Applicators and BARD¿ BrachySource¿ Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed up and sent to two hospitals. One Hospital ordered BrachySource¿ Iodine125 0.36 mCi radioactive seeds but instead received BrachySource¿ Iodine125 0.28 mCi radioactive seeds and vice versa for the other hospital affected.