2 — Class II medical device recall Z-1703-2026
Ongoing recall by Synthes (USA) Products LLC, reported 2026-04-08, distributed nationwide. A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68
| Recall number | Z-1703-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Synthes (USA) Products LLC, West Chester, PA, United States |
| Recall initiated | 2026-02-25 |
| Classified | 2026-03-30 |
| Reported by FDA | 2026-04-08 |
| Distributed | Nationwide (US) |
| Quantity | 33 units |
| UPC / GTIN / UDI-DI | 10886982335951 |
| Lot numbers | 68982P8 |
| Model numbers | 03.424.126-US |
Product
2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.
Reason for recall
A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.