Data Foundry

Medical device recalls 2018

Small Patient ECC Pack BEQ-TOP 22301 — Class II medical device recall Z-1707-2018

Terminated recall by Datascope Corporation, reported 2018-05-16, distributed in AL, AR, CA, DC, FL, IL, LA, MO…. The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. S

Recall numberZ-1707-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDatascope Corporation, Fairfield, NJ, United States
Recall initiated2017-10-19
Classified2018-05-08
Reported by FDA2018-05-16
Terminated2020-04-24
DistributedAL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, WA
Quantity1
Reason classespackaging
Lot numbers3000060036
Model numbers701049513

Product

Small Patient ECC Pack BEQ-TOP 22301, Catalog No. 701049513

Reason for recall

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1707-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.