Data Foundry

Medical device recalls 2015

Dimension Vista Systems LOCI Thyroid Stimulating Hormone — Class II medical device recall Z-1708-2015

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2015-06-10, distributed nationwide. Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results rangi

Recall numberZ-1708-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2015-04-20
Classified2015-06-04
Reported by FDA2015-06-10
Terminated2016-12-13
DistributedNationwide (US)
Countries outside the USAT, AU, BE, CA, CH, DE, DK, ES, FI, FR, GB, IT, JP, NL, NZ, PT, SK
Quantity2,256
Lot numbers14237AA
Model numbersK6412
Expiration dates2015-08-25

Product

Dimension Vista Systems LOCI Thyroid Stimulating Hormone; the TSH method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone in human serum and plasma on the Dimension Vista System.

Reason for recall

Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1708-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.