Data Foundry

Medical device recalls 2015

EVOTECH Endoscope Cleaner and Reprocessing System 208V — Class II medical device recall Z-1714-2015

Terminated recall by Advanced Sterilization Products, reported 2015-06-17, distributed nationwide. Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because

Recall numberZ-1714-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAdvanced Sterilization Products, Irvine, CA, United States
Recall initiated2015-05-11
Classified2015-06-05
Reported by FDA2015-06-17
Terminated2016-07-06
DistributedNationwide (US)
Countries outside the USAU, CA, SG, TH
Quantity825 units

Product

EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.

Reason for recall

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1714-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.