EVOTECH Endoscope Cleaner and Reprocessing System 220-230V — Class II medical device recall Z-1715-2015
Terminated recall by Advanced Sterilization Products, reported 2015-06-17, distributed nationwide. Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because
| Recall number | Z-1715-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Advanced Sterilization Products, Irvine, CA, United States |
| Recall initiated | 2015-05-11 |
| Classified | 2015-06-05 |
| Reported by FDA | 2015-06-17 |
| Terminated | 2016-07-06 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, SG, TH |
| Quantity | 44 units |
Product
EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.
Reason for recall
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1715-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.