BEQ-TOP 33700 ECC 3/8" PACK — Class II medical device recall Z-1716-2018

Terminated recall by Datascope Corporation, reported 2018-05-16, distributed in AL, AR, CA, DC, FL, IL, LA, MO…. The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. S

Recall numberZ-1716-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDatascope Corporation, Fairfield, NJ, United States
Recall initiated2017-10-19
Classified2018-05-08
Reported by FDA2018-05-16
Terminated2020-04-24
DistributedAL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, WA
Quantity22
Reason classespackaging
Lot numbers3000056811, 3000063949, 3000067795
Model numbers701053845

Product

BEQ-TOP 33700 ECC 3/8" PACK, Custom Tubing Pack with Bioline Coating, Catalog No. 701053845

Reason for recall

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.