Data Foundry

Medical device recalls 2019

OMNIBotics iBlock Cutting Guide — Class II medical device recall Z-1717-2019

Terminated recall by OMNIlife science Inc., reported 2019-06-12, distributed nationwide. The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during

Recall numberZ-1717-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOMNIlife science Inc., Raynham, MA, United States
Recall initiated2019-04-12
Classified2019-06-04
Reported by FDA2019-06-12
Terminated2020-05-14
DistributedNationwide (US)
Quantity9

Product

OMNIBotics iBlock Cutting Guide, 4146-1000 - Product Usage: The OMNIBotics iBlock Cutting Guide is a stainless-steel surgical instrument used during total knee arthroplasty. Specifically, the iBlock Cutting Guide is used in conjunction with the iBlock Saw Guide with the OMNIBot robotic Motor Unit, and for preparation of the bone for implant implantation. The instrument has been designed to facilitate bone preparation and implantation of the OMNI Apex Knee System.

Reason for recall

The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1717-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.