Data Foundry

Medical device recalls 2020

Sysmex CV-11 Sample Unit — Class II medical device recall Z-1717-2020

Terminated recall by Sysmex America, Inc., reported 2020-04-22, distributed in FL, IL, VT. When excessive amounts of urine spills on the rack transport track and reaches the electrical components of th

Recall numberZ-1717-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSysmex America, Inc., Mundelein, IL, United States
Recall initiated2020-01-21
Classified2020-04-15
Reported by FDA2020-04-22
Terminated2023-01-30
DistributedFL, IL, VT
Quantity40 units
Serial numbers11201, 11202, 12310, 12311, 12312, 12313, 12941, 12942, 12997, 13875, 14170, 14171, 14172, 14173, 14174, 14175, 14278, 14279, 14280, 14962, 14963, 14964, 14965, 14966, 14967 and 15 more

Product

Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.

Reason for recall

When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1717-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.