ICD-DR DDME3D4 MIRRO MRI — Class I medical device recall Z-1720-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1720-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 8,219 units |
| UPC / GTIN / UDI-DI | 00763000001582, 00763000001599, 00763000611941 |
| Serial numbers | 08708S, CWT6, CWT600002S, CWT600003S, CWT600005S, CWT600006S, CWT600007S, CWT600008S, CWT600009S, CWT600010S, CWT600011S, CWT600012S, CWT600013S, CWT600014S, CWT600016S, CWT600017S, CWT600018S, CWT600019S, CWT600020S, CWT600021S, CWT600022S, CWT600023S, CWT600024S, CWT600025S, CWT600026S and 475 more |
| Model numbers | DDME3D4 |
Product
ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.