Washer Disinfector Aquadis 56 — Class II medical device recall Z-1721-2025
Ongoing recall by Getinge Disinfection Ab, reported 2025-05-14, distributed nationwide. The frequency inverter is not programmed by supplier according to specification which causes the circulation p
| Recall number | Z-1721-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Getinge Disinfection Ab, Vaxjo, N/A, Sweden |
| Recall initiated | 2025-04-02 |
| Classified | 2025-05-05 |
| Reported by FDA | 2025-05-14 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BD, BH, BR, CA, CH, CN, CO, CR, CY, CZ, DE, DK, EC, EE, ES, FI, FR, GB, GR, HK, HR, HU, ID, IN, IQ, IR, IT, JP, KE, KR, KW, LT, LV, MX, MY, NG, NL, NO, OM, PH, PK, PL, RO, RS, SA, SE, SG, SI, SK, TH, TR, VN |
| Reason classes | packaging |
| UPC / GTIN / UDI-DI | 07340153710351 |
| Serial numbers | WAA105199, WAA105251, WAA105268, WAA105357, WAA105535, WAA105641, WAA105682, WAA105696, WAA105700, WAA105915, WAA105972, WAA105973, WAA106079, WAA106114, WAA106135, WAA106240, WAA106247, WAA106399, WAA106412, WAA106451, WAA106471, WAA106487, WAA106505, WAA106520, WAA106546 and 399 more |
| Model numbers | 56M |
Product
Washer Disinfector Aquadis 56; Model Number: 56M
Reason for recall
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.