Data Foundry

Medical device recalls 2014

Philips Medical Systems Computed Tomography X-ray systems — Class II medical device recall Z-1722-2014

Terminated recall by Philips Medical Systems (Cleveland) Inc, reported 2014-06-11, distributed nationwide. A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that

Recall numberZ-1722-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Medical Systems (Cleveland) Inc, Cleveland, OH, United States
Recall initiated2014-04-01
Classified2014-06-04
Reported by FDA2014-06-11
Terminated2018-06-26
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, DZ, EC, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VE, VN, ZA

Product

Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, and the MX8000 Dual v. EXP scanners)

Reason for recall

A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1722-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.