Data Foundry

Medical device recalls 2014

Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating… — Class II medical device recall Z-1726-2014

Terminated recall by Leica Microsystems, Inc., reported 2014-06-18, distributed in AZ, IL, MA, MO, NC, OR, TX, VA…. The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lo

Recall numberZ-1726-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmLeica Microsystems, Inc., Buffalo Grove, IL, United States
Recall initiated2013-09-11
Classified2014-06-06
Reported by FDA2014-06-18
Terminated2017-04-24
DistributedAZ, IL, MA, MO, NC, OR, TX, VA, WI
Countries outside the USAT, AU, CH, CL, CN, CO, CZ, ES, GB, GR, IL, IT, JP, KR, MY, NL, NO, OM, SA, SE
Quantity103 total

Product

Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Reason for recall

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1726-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.