GXL acetabular liners — Class II medical device recall Z-1728-2022

Ongoing recall by Exactech, Inc., reported 2022-09-21, distributed nationwide. Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device

Recall numberZ-1728-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmExactech, Inc., Gainesville, FL, United States
Recall initiated2022-08-11
Classified2022-09-09
Reported by FDA2022-09-21
DistributedNationwide (US)
Countries outside the USAR, BR, CL, CN, CO, EC, ES, GR, GT, IN, LB, PT
Quantity11,143 devices
UPC / GTIN / UDI-DI10885862013491, 10885862013507, 10885862013514, 10885862013606, 10885862013613, 10885862013682, 10885862013699, 10885862021984, 10885862021991, 10885862022004, 10885862022011, 10885862022028, 10885862022097, 10885862022103, 10885862022110, 10885862022202, 10885862022219, 10885862022226, 10885862023063, 10885862023070, 10885862023087, 10885862023094, 10885862023100, 10885862023117, 10885862023124 and 22 more
Serial numbers1007068, 1007069, 1007070, 1007071, 1007072, 1007073, 1007074, 1007075, 1007076, 1007077, 1007078, 1007079, 1007080, 1007081, 1007082, 1007083, 1007084, 1007085, 1007086, 1007087, 1007088, 1007089, 1007090, 1007091, 1007092 and 475 more
Model numbers104-28-43, 104-28-44, 104-28-45, 104-32-45, 104-32-46, 104-36-46, 104-36-47, 130-28-25, 130-28-26, 130-28-27, 130-28-28, 130-28-29, 130-32-27, 130-32-28, 130-32-29, 130-36-28, 130-36-29, 130-36-30, 132-28-24, 132-28-25, 132-28-26, 132-28-27, 132-28-28, 132-28-29, 132-28-30 and 23 more

Product

GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ F, REF 130-28-26 c. ACUMATCH GXL 0 DEGREE LINER 28MM SZ G, REF 130-28-27 d. ACUMATCH GXL 0 DEGREE LINER 28MM SZ H, REF 130-28-28 e. ACUMATCH GXL 0 DEGREE LINER 28MM SZ J, REF 130-28-29 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ G, REF 130-32-27 g. ACUMATCH GXL 0 DEGREE LINER 32MM SZ H, REF 130-32-28 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ J, REF 130-32-29 g. ACUMATCH GXL 0 DEGREE LINER 36MM SZ H, REF 130-36-28 h. ACUMATCH GXL 0 DEGREE LINER 36MM SZ J, REF 130-36-29 i. ACUMATCH GXL 0 DEGREE LINER 36MM SZ K, REF 130-36-30 j. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ E, REF 138-28-25 k. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ F, REF 138-28-26 l. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ G, REF 138-28-27 m. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ H, REF 138-28-28 n. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ J, REF 138-28-29 o. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ H, REF 138-36-28 p. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ J, REF 138-36-29 q. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ K, REF 138-36-30 r. ACUMATCH GXL 15 DEG +5 LAT LINER SZG 38MM, REF 138-36-27 s. ACUMATCH GXL 15DEG LINER 28MM SZ D, REF 132-28-24 t. ACUMATCH GXL 15DEG LINER 28MM SZ E, REF 132-28-25 u. ACUMATCH GXL 15DEG LINER 28MM SZ F, REF 132-28-26 v. ACUMATCH GXL 15DEG LINER 28MM SZ G, REF 132-28-27 w. ACUMATCH GXL 15DEG LINER 28MM SZ H, REF 132-28-28 x. ACUMATCH GXL 15DEG LINER 28MM SZ J, REF 132-28-29 y. ACUMATCH GXL 15DEG LINER 28MM SZ K, REF 132-28-30 z. ACUMATCH GXL 15DEG LINER 32MM SZ G, REF 132-32-27 aa. ACUMATCH GXL 15DEG LINER 32MM SZ H, REF 132-32-28 bb. ACUMATCH GXL 15DEG LINER 32MM SZ J, REF 132-32-29 cc. ACUMATCH GXL 15DEG LINER 32MM SZ K, REF 132-32-30 dd.ACUMATCH GXL 15DEG LINER 36MM SZ H, REF 132-36-28 ee. ACUMATCH GXL 15DEG LINER 36MM SZ J, REF 132-36-29 ff. ACUMATCH GXL 15DEG LINER 36MM SZ K, REF 132-36-30 gg. ACUMATCH GXL EXT COV LINER 28MM SZ E, REF 134-28-25 hh. ACUMATCH GXL EXT COV LINER 28MM SZ F, REF 134-28-26 ii. ACUMATCH GXL EXT COV LINER 28MM SZ G, REF 134-28-27 jj. ACUMATCH GXL EXT COV LINER 28MM SZ H, REF 134-28-28 kk. ACUMATCH GXL EXT COV LINER 28MM SZ J, REF 134-28-29 ll. ACUMATCH GXL EXT COV LINER 28MM SZ K, REF 134-28-30 mm. MCS GXL LINER 5/15 DEG 4448X28, REF 104-28-43 nn. MCS GXL LINER 5/15 DEG 5052X28, REF 104-28-44 oo. MCS GXL LINER 5/15 DEG 5458X28, REF 104-28-45 pp. MCS GXL LINER 5/15 DEG 5458X32, REF 104-32-45 qq. MCS GXL LINER 5/15 DEG 6064X32, REF 104-32-46 rr. MCS GXL LINER 5/15 DEG 6064X36, REF 104-36-46 ss. MCS GXL LINER 5/15 DEG 6670X36, REF 104-36-47

Reason for recall

Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.