Data Foundry

Medical device recalls 2014

HLD Systems Model 610 Washer/High Level Disinfector — Class II medical device recall Z-1729-2014

Terminated recall by Cenorin, LLC, reported 2014-06-18, distributed nationwide. Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential ri

Recall numberZ-1729-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCenorin, LLC, Kent, WA, United States
Recall initiated2014-04-24
Classified2014-06-06
Reported by FDA2014-06-18
Terminated2015-06-30
DistributedNationwide (US)
Quantity25 hld
Model numbers610

Product

HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.

Reason for recall

Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1729-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.